Please use this identifier to cite or link to this item:
http://nopr.niscpr.res.in/handle/123456789/14927| Title: | Studies on optimisation of collimator thickness and aperture for hot spot identification system using Monte Carlo technique |
| Authors: | Verma, Amit K Anilkumar, S Thakur, Vaishali M Jain, Amit Babu, D A R Sharma, D N |
| Keywords: | Hot spot identification system;Collimator thickness;Aperture |
| Issue Date: | Nov-2012 |
| Publisher: | NISCAIR-CSIR, India |
| Abstract: | The optimum collimator design in terms of
aperture and thickness for accomplishing improved performance for wide energy
radiation field has been investigated in the present paper. A 2" 2" NaI(Tl) detector was used as sensor in proposed Hot Spot Identification
system. The spectral response of the detector with various collimator apertures
and thickness was developed using Monte Carlo methods for optimizing the
collimator design. The optimum size of collimator aperture is found to be 2 mm
and thickness of 5 cm. Studies were also carried out to optimize lead shield
thickness around the detector.
|
| Page(s): | 874-875 |
| ISSN: | 0975-1041 (Online); 0019-5596 (Print) |
| Appears in Collections: | IJPAP Vol.50(11) [November 2012] |
Files in This Item:
| File | Description | Size | Format | |
|---|---|---|---|---|
| IJPAP 50(11) 874-875.pdf | 75.09 kB | Adobe PDF | View/Open |
Items in NOPR are protected by copyright, with all rights reserved, unless otherwise indicated.
2" NaI(Tl) detector was used as sensor in proposed Hot Spot Identification
system. The spectral response of the detector with various collimator apertures
and thickness was developed using Monte Carlo methods for optimizing the
collimator design. The optimum size of collimator aperture is found to be 2 mm
and thickness of 5 cm. Studies were also carried out to optimize lead shield
thickness around the detector.