Please use this identifier to cite or link to this item: http://nopr.niscpr.res.in/handle/123456789/16735
Title: Gas chromatographic retention indices for the N-methyl-N-substituted benzamides on capillary columns
Authors: Antonovic, Dusan G
Mijin, Dusan Z
Petrovic, Slobodan D
Stojanovic, Nadezda D
Nikolic, Aleksandar D
Issue Date: Sep-2000
Publisher: NISCAIR-CSIR, India
Abstract: Gas chromatographic retention indices (1) for series of 20 synthesized N-methyl-N-substituted benzamides on DB-1,DB-5 and DB-WAX capillary columns isothermal at 170, 190 and 210oC , and using temperature programmed gas chromatography (TPGC) have been determined within standard deviation of 0.6 i.u. The isothermal data obtained on these columns and TPGC data are compared and a comparison of the retention indices, ΔI and di/dT values presented as linear regression equations with correlation coefficient greater than 0.9970 within standard deviation of 0.6 i.u. The best resolution of the structural isomers is obtained using DB-1 capillary column.
Page(s): 694-699
ISSN: 0975-0983(Online); 0376-4699(Print)
Appears in Collections:IJC-B Vol.39B(09) [September 2000]

Files in This Item:
File Description SizeFormat 
IJCB 39B(9) 694-699.pdf1.39 MBAdobe PDFView/Open