http://nopr.niscpr.res.in/handle/123456789/17990| Title: | External proton-induced X-ray emission: A simple method for multi-elemental analysis of water |
| Authors: | Nayak, P K Rautray, T R Vijayan, V |
| Issue Date: | Oct-2006 |
| Publisher: | NISCAIR-CSIR, India |
| IPC Code: | Int. Cl.8 G01 N23/221 |
| Abstract: | An accelerator based rapid method for multi-elemental analysis of water samples has been presented using pre-concentration (metal ion as carbamates) technique followed by the use of external PIXE set-up. The problems regarding time required in creating vacuum and the irradiation of number of specimens (usually four to six numbers in a single run) in conventional vacuum-PIXE technique has been successfully overcome. The method is expected to be useful for multi-elemental analysis of various liquid samples on routine basis. |
| Page(s): | 2233-2237 |
| ISSN: | 0975-0975(Online); 0376-4710(Print) |