Please use this identifier to cite or link to this item:
http://nopr.niscpr.res.in/handle/123456789/20554| Title: | Instrumentation for a retarding potential analyzer for measurement of electron parameters |
| Authors: | Rajput, S S Garg, S C |
| Issue Date: | Oct-2002 |
| Publisher: | NISCAIR-CSIR, India |
| Abstract: | Various design considerations and their implementation in an instrument, designed for measurement of electron parameters on-board Indian satellites SROSS-C2, along its orbital path are presented. The proposed instruments measures the electron density in the range 5x102 - 5x106 cm-3 electron temperatures from 500 K to 4000 K, and photoelectron flux of energy 3-30 eV of the order 108 cm- s-1 with full -scale instrument sensitivity 0.2%. Experimental results obtained during testing and flight demonstrate the instrument performance. |
| Page(s): | 315-322 |
| ISSN: | 0975-1017 (Online); 0971-4588 (Print) |
| Appears in Collections: | IJEMS Vol.09(5) [October 2002] |