Please use this identifier to cite or link to this item: http://nopr.niscpr.res.in/handle/123456789/27033
Full metadata record
DC FieldValueLanguage
dc.contributor.authorSingh, Ajeet-
dc.contributor.authorSharma, Sudhir Kumar-
dc.contributor.authorJaiswal, Shiv Kumar-
dc.contributor.authorOjha, Vijay Narain-
dc.date.accessioned2014-02-27T07:25:24Z-
dc.date.available2014-02-27T07:25:24Z-
dc.date.issued2005-06-
dc.identifier.issn0975-2412 (Online); 0771-7706 (Print)-
dc.identifier.urihttp://hdl.handle.net/123456789/27033-
dc.description74-77en_US
dc.description.abstract    In computer age, there is an important contribution of automation in improving the error free precision calibration and to save the time. Automatic measurement is normally done using computer in high precision system and calibration.
    In the present paper, we discuss in detail about the automatic process and as a case study the precision calibration of digital multi meter by high precision calibrator traceable to ‘‘National Standard’’. The commercial software MET/CAL was used in the automation. We also discuss the process of the validation of this software prior to its use. The uncertainty evaluation reported in this paper is as per the ISO ‘‘GUM’’ document.
en_US
dc.language.isoen_USen_US
dc.publisherNISCAIR-CSIR, Indiaen_US
dc.rights CC Attribution-Noncommercial-No Derivative Works 2.5 Indiaen_US
dc.sourceBVAAP Vol.13(1) [June 2005]en_US
dc.titleApplication of Automation in Calibration of Multimeter using Precision Calibratoren_US
dc.typeArticleen_US
Appears in Collections:BVAAP Vol.13(1) [June 2005]

Files in This Item:
File Description SizeFormat 
BVAAP 13(1) 74-77.pdf652.11 kBAdobe PDFView/Open


Items in NOPR are protected by copyright, with all rights reserved, unless otherwise indicated.