Please use this identifier to cite or link to this item: http://nopr.niscpr.res.in/handle/123456789/2846
Full metadata record
DC FieldValueLanguage
dc.contributor.authorSharma, P-
dc.contributor.authorSongara, S-
dc.date.accessioned2009-01-22T06:40:38Z-
dc.date.available2009-01-22T06:40:38Z-
dc.date.issued2008-09-
dc.identifier.issn0971–457X-
dc.identifier.urihttp://hdl.handle.net/123456789/2846-
dc.description504-506en_US
dc.description.abstractA simple, rapid and sensitive differential pulse polarographic (DPP) method for the determination of low concentration of chlorate in water samples is presented. It is based on the measurement of increased current of titanium (IV) polarographic reduction wave by the addition of chlorate in oxalic acid medium. A linearity between 1.0 μg mL⁻¹ and 41.20 μg mL⁻¹ of chlorate concentration was obtained with a correlation of coefficient of 0.99.en_US
dc.language.isoen_USen_US
dc.publisherCSIRen_US
dc.sourceIJCT Vol.15(5) [September 2008]en_US
dc.subjectChlorate determinationen_US
dc.subjectDifferential pulse polarographyen_US
dc.subjectWater samplesen_US
dc.titleVoltammetric trace determination of sub- μg level chlorate in natural watersen_US
dc.typeArticleen_US
Appears in Collections:IJCT Vol.15(5) [September 2008]

Files in This Item:
File Description SizeFormat 
IJCT 15(5) 504-506.pdf73.19 kBAdobe PDFView/Open