Please use this identifier to cite or link to this item: http://nopr.niscpr.res.in/handle/123456789/29979
Full metadata record
DC FieldValueLanguage
dc.contributor.authorGeorgiuss, R S-
dc.date.accessioned2014-11-13T09:47:14Z-
dc.date.available2014-11-13T09:47:14Z-
dc.date.issued1994-12-
dc.identifier.issn0975-1017 (Online); 0971-4588 (Print)-
dc.identifier.urihttp://hdl.handle.net/123456789/29979-
dc.description305-308en_US
dc.description.abstractAn investigation of freezing and melting temperature of two samples of lead (purity> 99.99%) has been made. Using the induced freezing technique, plateaus of essentially constant (±0.0002o C) temperature with duration of3 h are obtained on the cooling curves of these samples. Alloy melting ranges have been measured following different types of freezing with overnight anneals near the solidus temperature. These parameters are found to be useful in the selection of the samples of the highest purity and at the same time showed that an uncertainty of 0.004o C above the value of the liquid us point may exist.en_US
dc.language.isoen_USen_US
dc.publisherNISCAIR-CSIR, Indiaen_US
dc.rights CC Attribution-Noncommercial-No Derivative Works 2.5 Indiaen_US
dc.sourceIJEMS Vol.01(6) [December 1994]en_US
dc.titleFreezing points of high purity lead as precision temperature standardsen_US
dc.typeArticleen_US
Appears in Collections:IJEMS Vol.01(6) [December 1994]

Files in This Item:
File Description SizeFormat 
IJEMS 1(6) 305-308.pdf253.56 kBAdobe PDFView/Open