Please use this identifier to cite or link to this item: http://nopr.niscpr.res.in/handle/123456789/30823
Full metadata record
DC FieldValueLanguage
dc.contributor.authorAmarchand, S-
dc.contributor.authorMenon, S K-
dc.contributor.authorAgrawal, Y K-
dc.date.accessioned2015-03-11T09:22:32Z-
dc.date.available2015-03-11T09:22:32Z-
dc.date.issued1998-03-
dc.identifier.issn0975-0991 (Online); 0971-457X (Print)-
dc.identifier.urihttp://hdl.handle.net/123456789/30823-
dc.description99-103en_US
dc.description.abstractAn ion selective electrode used in the direct determination of water hardness in tap-water has been fabricated using Mg (II)-N-phenylbenzohydroxamate complex in the membrane. The electrode characteristics exhibited almost equal selectivities for Ca2+ and Mg2+ which enabled the simultaneous determination of both the ions. The results have been compared with EDTA titrimetry.en_US
dc.language.isoen_USen_US
dc.publisherNISCAIR-CSIR, Indiaen_US
dc.rights CC Attribution-Noncommercial-No Derivative Works 2.5 Indiaen_US
dc.sourceIJCT Vol.05(2) [March 1998]en_US
dc.titleWater hardness determination using Mg(II) ion selective electrodeen_US
dc.typeArticleen_US
Appears in Collections:IJCT Vol.05(2) [March 1998]