Please use this identifier to cite or link to this item:
http://nopr.niscpr.res.in/handle/123456789/32344| Title: | Effect of sample containers on the EDXRF determination of uranium in aqueous solutions |
| Authors: | Natarajan, V Hon, N S Bhide, M K Godbole, S V |
| Keywords: | EDXRF;Glass containers;Plastic containers;Uranium |
| Issue Date: | Jan-2015 |
| Publisher: | NISCAIR-CSIR, India |
| Abstract: | The utility of plastic/glass containers in place of
regular sample cups made of high density polyethylene for the EDXRF
determination of uranium has been examined. A series of uranium solution
standards have been prepared and calibrated using
different containers. For U solution in plastic cups with thin mylar film base,
linearity between U L X-ray
intensity and its concentration is observed upto U 20 mg/mL; for higher concentrations, non-linear trend is observed, indicating
matrix absorption effects. For U
solution in plastic vials, similar behaviour is observed, with reduced matrix
absorption. In the case of U solutions
in sealed plastic vials, the reduction in the intensities of U X-rays of standard
solutions is about 30-70% of those for U in unsealed vials. In the case of
glass vials, the intensity of U X-ray peak decreases drastically as compared to
that for U in plastic cups with thin mylar film base. The present study suggest
the utility of sealed plastic vials for the fast preliminary determination of
uranium in low active process / laboratory waste solutions with a precision of
about 5-15%.
|
| Page(s): | 82-84 |
| ISSN: | 0975-0991 (Online); 0971-457X (Print) |
| Appears in Collections: | IJCT Vol.22(1-2) [January-March 2015] |
Files in This Item:
| File | Description | Size | Format | |
|---|---|---|---|---|
| IJCT 22(1-2) 82-84.pdf | 73.98 kB | Adobe PDF | View/Open |
Items in NOPR are protected by copyright, with all rights reserved, unless otherwise indicated.
X-ray
intensity and its concentration is observed upto U
20 mg/mL; for higher concentrations, non-linear trend is observed, indicating
matrix absorption effects. For U
solution in plastic vials, similar behaviour is observed, with reduced matrix
absorption. In the case of U solutions
in sealed plastic vials, the reduction in the intensities of U X-rays of standard
solutions is about 30-70% of those for U in unsealed vials. In the case of
glass vials, the intensity of U X-ray peak decreases drastically as compared to
that for U in plastic cups with thin mylar film base. The present study suggest
the utility of sealed plastic vials for the fast preliminary determination of
uranium in low active process / laboratory waste solutions with a precision of
about 5-15%.