Title: Effect of sample containers on the EDXRF determination of uranium in aqueous solutions
Authors: Natarajan, V
Hon, N S
Bhide, M K
Godbole, S V
Keywords: EDXRF;Glass containers;Plastic containers;Uranium
Issue Date: Jan-2015
Publisher: NISCAIR-CSIR, India
Abstract: The utility of plastic/glass containers in place of regular sample cups made of high density polyethylene for the EDXRF determination of uranium has been examined. A series of uranium solution standards have been prepared and calibrated using different containers. For U solution in plastic cups with thin mylar film base, linearity between U L X-ray intensity and its concentration is observed upto U 20 mg/mL; for higher concentrations, non-linear trend is observed, indicating matrix absorption effects.  For U solution in plastic vials, similar behaviour is observed, with reduced matrix absorption.  In the case of U solutions in sealed plastic vials, the reduction in the intensities of U X-rays of standard solutions is about 30-70% of those for U in unsealed vials. In the case of glass vials, the intensity of U X-ray peak decreases drastically as compared to that for U in plastic cups with thin mylar film base. The present study suggest the utility of sealed plastic vials for the fast preliminary determination of uranium in low active process / laboratory waste solutions with a precision of about 5-15%.
Page(s): 82-84
ISSN: 0975-0991 (Online); 0971-457X (Print)
Appears in Collections:IJCT Vol.22(1-2) [January-March 2015]

Files in This Item:
File Description SizeFormat 
IJCT 22(1-2) 82-84.pdf73.98 kBAdobe PDFView/Open


Items in NOPR are protected by copyright, with all rights reserved, unless otherwise indicated.