Please use this identifier to cite or link to this item: http://nopr.niscpr.res.in/handle/123456789/36817
Full metadata record
DC FieldValueLanguage
dc.contributor.authorVerma, S D-
dc.contributor.authorSinha, Vijaya-
dc.date.accessioned2016-11-07T07:25:06Z-
dc.date.available2016-11-07T07:25:06Z-
dc.date.issued1982-10-
dc.identifier.issn0975-105X (Online); 0367-8393 (Print)-
dc.identifier.urihttp://nopr.niscair.res.in/handle/123456789/36817-
dc.description205-208en_US
dc.description.abstractA silicon surface barrier detector is designed and fabricated to observe IX-particles from various radioactive sources. The energy spectrum of monoenergetic α–particles from 2&1 Am source using this detector is obtained with the help of a 256 channel microprocessor controlled pulse height analyzer (MCA). A shift in the peak of the energy spectrum towards the lower channel is observed if the pressure between the source and the detector is increased. This can be used to observe the pressure. A differential counting method is described to determine the pressure accurately. The application of this device in a balloon flight experiment is discussed.en_US
dc.language.isoen_USen_US
dc.publisherNISCAIR-CSIR, Indiaen_US
dc.rights CC Attribution-Noncommercial-No Derivative Works 2.5 Indiaen_US
dc.sourceIJRSP Vol.11(5) [October 1982]en_US
dc.titleSensitive Pressure Gauge Using Silicon Surface Barrier Detectoren_US
dc.typeArticleen_US
Appears in Collections:IJRSP Vol.11(5) [October 1982]

Files in This Item:
File Description SizeFormat 
IJRSP 11(5) 205-208.pdf239.79 kBAdobe PDFView/Open