Please use this identifier to cite or link to this item: http://nopr.niscpr.res.in/handle/123456789/43947
Title: An amperometric sensor for -ascorbic acid based on graphite paste electrode
Authors: Pandey, P C
Pandey, V
Mehta, S
Issue Date: Aug-1993
Publisher: NISCAIR-CSIR, India
Abstract: An amperometric sensor for -ascorbic acid based on graphite paste electrode with and without incorporating tetracyanoquinodimethane (TCNQ) is described. DC cyclic voltammetry indicates that -ascorbic acid can be oxidized at very low potential (0 mV vs SCE). DC cyclic voltammetry of TCNQ modified graphite paste electrode shows two anodic peaks for the oxidation of -ascorbic acid due to diffusion current and catalytic current. The catalytic wave is obtained due to the mediated mechanism of electron exchange through TCNQ. The second order rate constant (0.14 x 104 1 mol-1 s-1) for the reaction between -ascorbic acid and TCNQ has also been calculated. The TCNQ modified electrode shows long term stability.
Page(s): 701-703
ISSN: 0975-0975(Online); 0376-4710(Print)
Appears in Collections:IJC-A Vol.32A(08) [August 1993]

Files in This Item:
File Description SizeFormat 
IJCA 32A(8) 701-703.pdf193.44 kBAdobe PDFView/Open


Items in NOPR are protected by copyright, with all rights reserved, unless otherwise indicated.