dc.description761-766en_US
dc.description.abstractAn inexpensive controlled drug infusion system based around 80C85 microprocessor with RAM, EPROM and peripheral interface devices is developed. In this system, the drop rate selection is done by BCD thumb wheel switches from 1 to 99 drops/min, LCD display is used for displaying the parameters settings, number of drops infused and various alarms messages for patient safety and unique drop sensing assembly is incorporated for accurate detection of falling drops in drip chamber. The prototype has been tested successfully on neonates.en_US
dc.language.isoen_USen_US
dc.publisherCSIRen_US
dc.relation.ispartofseriesA61M5/14en_US
dc.sourceJSIR Vol.64(10) [October 2005]en_US
dc.subjectDrug infusionen_US
dc.subjectDrop sensing assemblyen_US
dc.subjectMicroprocessoren_US
dc.subjectAlarmsen_US
dc.subjectI V seten_US
dc.titleDesign and implementation of controlled drug infusion systemen_US