Please use this identifier to cite or link to this item: http://nopr.niscpr.res.in/handle/123456789/62863
metadata.dc.identifier.doi: https://doi.org/10.56042/jsir.v82i11.5506
Title: Comparative Analysis of Porous Titanium Spinal Cage with Conventional Spinal Cages: A Finite Element Study
Authors: Kumar, Prashant
Bhardwaj, Rahul
Matharu, Amrit Lal
Meena, Vijay Kumar
Keywords: Additive manufacturing;CFR-PEEK;PEEK;PLIF;Ti6Al4V ELI
Issue Date: Nov-2023
Publisher: NIScPR-CSIR,India
Abstract: The objective of this study is to compare the stress shielding effect of various conventional as well as modified additive manufactured porous materials used for spinal cages. A finite element study was performed by changing the design (fully porous and hybrid) and the materials (PEEK, CFR-PEEK, Titanium) of spinal cages. All the models were simulated under uniaxial compression, to study the stress shielding effect. The Finite Element Analysis results showed that the hybrid spinal cage transfers more stress to its adjacent vertebrae than the other design configurations under uniaxial compression. The hybrid titanium cage was most effective in reducing the stress shielding effect. The hybrid cage is stronger than PEEK & CFR-PEEK cages, however, due to the porous structure reduced stress shielding was observed.
Page(s): 1134-1142
ISSN: 0022-4456 (Print); 0975-1084 (Online)
Appears in Collections:JSIR Vol.82(11) [November 2023]

Files in This Item:
File Description SizeFormat 
JSIR 82(11) 1134-1142.pdf2.85 MBAdobe PDFView/Open