Please use this identifier to cite or link to this item:
http://nopr.niscpr.res.in/handle/123456789/7039Full metadata record
| DC Field | Value | Language |
|---|---|---|
| dc.contributor.author | Yadav, Sanjay | - |
| dc.contributor.author | Gupta, V K | - |
| dc.contributor.author | Bandyopadhyay, A K | - |
| dc.date.accessioned | 2010-01-01T11:29:28Z | - |
| dc.date.available | 2010-01-01T11:29:28Z | - |
| dc.date.issued | 2010-01 | - |
| dc.identifier.issn | 0975-1084 (Online); 0022-4456 (Print) | - |
| dc.identifier.uri | http://hdl.handle.net/123456789/7039 | - |
| dc.description | 27-33 | en_US |
| dc.description.abstract | Proficiency testing of pressure measurement was carried out among 21 laboratories using digital pressure calibrator. Total 178 pressure measurements (pressure range, 7-70 MPa) were recorded at 10 equally spaced pressure points. Normalized error (En) values of 15 laboratories were within acceptable limits as per guidelines given by National Measurement Institute, Govt of India. En values of 3 laboratories were found beyond the acceptable limit. | en_US |
| dc.language.iso | en_US | en_US |
| dc.publisher | CSIR | en_US |
| dc.source | JSIR Vol.69(01) [January 2010] | en_US |
| dc.subject | Interlaboratory comparison | en_US |
| dc.subject | Pressure calibration | en_US |
| dc.subject | Pressure metrology | en_US |
| dc.subject | Proficiency testing | en_US |
| dc.title | Standardization of pressure calibration (7-70 MPa) using digital pressure calibrator | en_US |
| dc.type | Article | en_US |
| Appears in Collections: | JSIR Vol.69(01) [January 2010] | |
Files in This Item:
| File | Description | Size | Format | |
|---|---|---|---|---|
| JSIR 69(1) 27-33.pdf | 114.35 kB | Adobe PDF | View/Open |
Items in NOPR are protected by copyright, with all rights reserved, unless otherwise indicated.