Please use this identifier to cite or link to this item: http://nopr.niscpr.res.in/handle/123456789/915
Full metadata record
DC FieldValueLanguage
dc.contributor.authorNanda, D-
dc.contributor.authorOak, M S-
dc.contributor.authorKumar, M Pravin-
dc.date.accessioned2008-04-10T09:16:27Z-
dc.date.available2008-04-10T09:16:27Z-
dc.date.issued2007-02-
dc.identifier.urihttp://hdl.handle.net/123456789/915-
dc.description258-262en_US
dc.description.abstractA poly(vinyl chloride) based membrane ion selective electrode has been developed in which acrylic acid has been grafted on poly(vinyl chloride) powder. The ion selective electrode system, prepared without an internal solution, contains DBP (plasticizer) and NaTPB (anion excluder), and shows near Nernstian response to Mg²⁺, Cu²⁺ and super-Nernstian response to Ca²⁺ in the concentration range 10⁻¹ – 10-5 mol L⁻¹. EDTA has been used to mask the interference of transition metal ions. The best performance of the electrode was obtained when it was stored in 10⁻² mol/L Ca²⁺ solution when the electrode was not in use. The electrode, with response time ≤ 20 s, has been found useful in the determination of concentration of Ca²⁺ and hardness in river water.en_US
dc.language.isoen_USen_US
dc.publisherCSIRen_US
dc.relation.ispartofseriesInt. Cl.⁸ G01N27/333en_US
dc.sourceIJCA Vol.46A(2) [February 2007]en_US
dc.titleAcrylic acid grafted PVC membrane based ion selective electrode for calcium and hardness measurement of wateren_US
dc.typeArticleen_US
Appears in Collections:IJC-A Vol.46A(02) [February 2007]