Please use this identifier to cite or link to this item: http://nopr.niscpr.res.in/handle/123456789/30088
Title: Quantitative determination of ciprofloxacin residues for cleaning validation in manufacturing equipment surface
Authors: Shah, Purvi
Champaneri, Naitik
Solanki, Kamlesh
Thakkar, Vaishali
Patel, Kalpana
Gandhi, Tejal
Keywords: Ciprofloxacin;Cleaning validation;Manufacturing equipment surfaces;Total organic carbon Vcsh analysis;UV spectroscopic method
Issue Date: Jul-2014
Publisher: NISCAIR-CSIR, India
Abstract: In this study, UV spectroscopic and total organic carbon (TOC) Vcsh methods have been developed and validated for the determination of trace amount of ciprofloxacin residues on the manufacturing equipment surfaces and to confirm the efficiency of the cleaning procedure. Ciprofloxacin has been selected on the basis of worst case rating approach. The established limit of contamination value for ciprofloxacin is found to be 36.27 µg that should not be exceeded during consecutive equipment cleaning trials. Ciprofloxacin exhibits good linearity in the range 2-10 µg/mL for UV spectroscopic method and 0.25-1.5 µg/mL for TOC Vcsh analysis. Percentage recovery from spiked stainless steel plate using swab sampling technique is found in the limit of 90.74% and 91.19 % in UV spectroscopic and TOC Vcsh method respectively. Both methods are simple, cheap and less time consuming as well as suitable for quantitative determination of ciprofloxacin on manufacturing equipment surfaces well below the limit of contamination. The validated methods are found to be satisfactory for the demonstration of cleaning validation of ciprofloxacin residue on manufacturing equipment surface.
Page(s): 233-237
ISSN: 0975-0991 (Online); 0971-457X (Print)
Appears in Collections:IJCT Vol.21(4) [July 2014]

Files in This Item:
File Description SizeFormat 
IJCT 21(4) 233-237.pdf192.59 kBAdobe PDFView/Open


Items in NOPR are protected by copyright, with all rights reserved, unless otherwise indicated.